Volume 39, Issue 11 , Pages 1120-1126, November 2010
Biomechanical three-dimensional finite-element analysis of maxillary prostheses with implants. Design of number and position of implants for maxillary prostheses after hemimaxillectomy
Abstract
The present study analyzed stress distributions in craniofacial structures around implant-supported maxillary prostheses. Using post-hemimaxillectomy computed tomography (CT) of a patient, the authors constructed a three-dimensional (3D) solid model using Digital Imaging and Communications in Medicine data (DICOM data) for maxillofacial and cranial bones. The effects of different prosthesis designs on stress distributions in craniofacial bones and osseous tissues around the implants were biomechanically investigated using 3D finite-element analysis. Maxillary prostheses were designed with 2 implants in the zygoma on the affected side and 2–3 implants in the maxillary alveolar bone on the unaffected side, without using a cantilever. Zygomatic implants provided suitable stress dispersal to the zygomatic and craniofacial bones on the affected side. This information is useful for designing maxillary prostheses.
Key words: zygomatic implants, maxillectomy, digital biomechanical model of the craniofacial skeleton, three-dimensional finite-element analysis, stress distribution
To access this article, please choose from the options below
PII: S0901-5027(10)00290-0
doi:10.1016/j.ijom.2010.06.011
© 2010 International Association of Oral and Maxillofacial Surgeons. Published by Elsevier Inc. All rights reserved.